Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29798 |
| Brand Name: | Topical Transducer |
| Model: | Philips EPIQ Ultrasound Systems with versions 9.0, 9.0.1, and 9.0.2 which, when used together (X5-1 transducer) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Ultrasound LLC |
| Brand Name: | General Ultrasound Imaging System |
| Model: | Philips EPIQ Ultrasound Systems with versions 9.0, 9.0.1, and 9.0.2 which, when used together (Ultrasound System) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Ultrasound LLC |
| Issue: | Update information: If the X5-1c transducer is connected to the right-most port (when viewing from the front), this problem will not occur. Please ensure your X5-1c is connected to the right-most connector until the software update (FCO79500561) has been installed |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/8/2022 |