Detail of Recall

Type of Product:Device
Medsafe Reference:29769
Brand Name:Automated Endoscope Reprocessor
Model:Getinge ED-Flow AER (Automated Endoscope Reprocessor); and Getinge ED-Flow SD AER - multiple serial numbers
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Getinge Lancer
Issue:The manufacturer has discovered that a temperature sensor in the ED-FLOW AER chamber may not work as intended and may provide incorrect temperature readings for washing and disinfection temperature during process. The result on the process ticket has shown OK despite the temperature not having reached the parameters required for validated biocidal activity.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/8/2022