Type of Product:Device
Medsafe Reference:29762
Brand Name:CN - Percept PC BrainSense B35200
Model:A610 Clinician Software Application
Affected:
Software version:3.0.1048, 3.0.1057, 3.0.1062, and 3.0.1081
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:The manufacturer has identified an anomaly with the clinician programmer application (A610) when interrogating an Activa™ PC or Activa™ RC.
In rare instances, this anomaly may present during a follow-up session if the patient was previously programmed with an A610 DBS Clinician Programmer Application Version v.2.0.x (2.0.4584, 2.0.4594, 2.0.4605, and 2.0.4630) and later that same Activa™ PC or Activa™ RC patient's device is interrogated with a clinician programmer tablet running A610 3.0.x
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/7/2022