Detail of Recall

Type of Product:Device
Medsafe Reference:29727
Brand Name:Suzhou DXR MDR Systems
Model:Philips DigitalDiagnost C50 1.1
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Healthcare (Suzhou) Co Ltd
Brand Name:DuraDiagnost, DigitalDiagnost C50
Model:Philips DigitalDiagnost C50
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Healthcare (Suzhou) Co Ltd
Issue:The manufacturer has become aware of an anomaly with our Philips DigitalDiagnost C50 when following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The anatomic position markers become malpositioned and could potentially be associated with the opposite side of the anatomy.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/7/2022