Detail of Recall

Type of Product:Device
Medsafe Reference:29725
Brand Name:BD Veritor™ Plus Analyzer
Model:
Affected:Batch numbers - JB202206, JB202211, JB202310, JB202312, JB202312, JB202403
Software version:
Recalling Organisation:Becton Dickinson Pty Ltd
80 Rushdale Street
Knoxfield
VIC 3180
Australia
Contact Information:
Manufacturer:Becton Dickinson and Company (USA)
Issue:The manufacturer has received reporting that the BD Veritor Plus Analyzer may have the potential to overheat and/or cause fire. The initial investigation found that the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD with the BD Veritor Plus Analyzer, is used to power the instrument.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/7/2022