Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29719
Brand Name:HLS
Model:BE-HLS 7050 (SHLS Set Advanced 7.0), BE-HLS 5050 (SHLS Set Advanced 5.0), BO-HLS 7050 (S/HIT Set Advanced 7.0), BEQ-HLS 7050 USA (SHLS Set Advanced 7.0), BEQ-HLS 5050 USA (SHLS Set Advanced 5.0), BEQ-HLS 7050-CA (SHLS Set Advanced 7.0),
Affected:Additional models: BEQ-HLS 5050-CA (SHLS Set Advanced 5.0), BE-HLS 7050 Japan (SHLS Set Advanced 7.0).
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:The manufacturer has received customer complaints of damage to the primary packaging of the HLS Set Advanced. In all customer complaints it was reported that the Tyvek cover was damaged. The appearance of the packaging failure can be hole(s) in the Tyvek. The damages are caused by the tabs of the venous probe holder within the packaging.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/7/2023