Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29717
Brand Name:Minisart NML Syringe Filters
Model:16555-K
Affected:Multiple batch numbers
Software version:
Recalling Organisation:Micronz Limited
40 Naples Street
Martinborough 5711
Contact Information:
Manufacturer:Sartorius Stedim Biotech GmbH
Brand Name:Minisart NML Syringe Filters
Model:16534-K
Affected:Multiple batch numbers
Software version:
Recalling Organisation:Micronz Limited
40 Naples Street
Martinborough 5711
Contact Information:
Manufacturer:Sartorius Stedim Biotech GmbH
Issue:The instructions for use will be updated to include a specification limitation regarding the maximum pressure (not to exceed 1 bar) while using filters 16596-HYK and 17575-ACK in the OUT-IN direction.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/7/2022