Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29715
Brand Name:NMD - Vanta™ with AdaptiveStim™ Neurostimulator System
Model:A71200: Vanta Clinician Programmer Application (CP App)
Affected:
Software version:v2.0.2455
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:As previously communicated by the manufacturer, in July 2022, in these rare occurrences during initial programming, the Vanta CP App will display the Error Code 1502 message because the Vanta INS Reset Block ID log is full. Once the log is full, all subsequent interrogations with the affected device will not be possible. So the manufactuere performed an INS diagnostic and log reset, which enabled the Vanta INSs to be interrogated by the Vanta CP App and complete programming.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/12/2022