Type of Product:Device
Medsafe Reference:29704
Brand Name:Medicina Oral tip syringes
Model:
Affected:(OT01) HM18147, HM19025, HM19088, HM19164, HM20063, HM21015, HM21072. (OT05) HM22028. (OT25) HM20064, HM21061
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Medicina Ltd
Brand Name:ENFit Enteral Syringes
Model:
Affected:(LHE01LD): HM19016, HM19083, HM20005, HM20054, HM21044, HM22018. (LHE05): HM18010, HM18052, HM18113, HM19018, HM19061, HM19084, HM19146, HM20006, HM20032, HM20055, HM20095, HM21037, HM21069, HM22019. (9LHE25LD): HM20031, HM20049, HM21033, HM22008.
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Medicina Ltd
Issue:The manufacturer has discovered that the plunger component for the ENFit Enteral Syringes has been manufactured using polyethylene, while Medicina's technical documents state that the material should be polypropylene.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/7/2022