Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29704 |
| Brand Name: | Medicina Oral tip syringes |
| Model: | |
| Affected: | (OT01) HM18147, HM19025, HM19088, HM19164, HM20063, HM21015, HM21072. (OT05) HM22028. (OT25) HM20064, HM21061 |
| Software version: | |
| Recalling Organisation: | New Zealand Medical & Scientific Ltd Level 4, Building 10 666 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 09 259 4062 |
| Manufacturer: | Medicina Ltd |
| Brand Name: | ENFit Enteral Syringes |
| Model: | |
| Affected: | (LHE01LD): HM19016, HM19083, HM20005, HM20054, HM21044, HM22018. (LHE05): HM18010, HM18052, HM18113, HM19018, HM19061, HM19084, HM19146, HM20006, HM20032, HM20055, HM20095, HM21037, HM21069, HM22019. (9LHE25LD): HM20031, HM20049, HM21033, HM22008. |
| Software version: | |
| Recalling Organisation: | New Zealand Medical & Scientific Ltd Level 4, Building 10 666 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 09 259 4062 |
| Manufacturer: | Medicina Ltd |
| Issue: | The manufacturer has discovered that the plunger component for the ENFit Enteral Syringes has been manufactured using polyethylene, while Medicina's technical documents state that the material should be polypropylene. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/7/2022 |