Detail of Recall

Type of Product:Device
Medsafe Reference:29667
Brand Name:Hamilton/47244/Ventilators
Model:
Affected:Loose status indicator board (Part Number 160021)
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hamilton Medical AG
Issue:The manufacturer has identified a hardware issue related to the status indicator board which can become loose and may result in a loose contact to the mainboard and potentially oxidation of this contact due to ingress of water (e.g., disinfectants).
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/7/2022