Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29647
Brand Name:DIre_Orthopaedic prosthesis implantation instrument [13180]
Model:
Affected:Lot number - AB4810808
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:DePuy Ireland
Issue:The manufacturer has received communications from some customers indicating that the spring in tthe DIre Orthopaedic prosthesis implantation instrument had corroded. Upon further investigation, they discovered that nine (9) lots of the subject product had been manufactured with the incorrect grade of steel. The incorrect steel spring material has the potential to oxidize, leading to corrosion. All nine (9) lots of the subject product, regardless of whether the spring has oxidized, are covered by this recall.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2022