Affected:
Software version:2868 TY+ software
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Issue:The manufacturer has identified that the Model 2868 TY+ software on the Model 3120 ZOOM programmer has been uninstalled or is outdated. As such, users may observe unexpected behaviours when attempting communications with transvenous defibrillators2 including the potential inability to interrogate or the display of incorrect/missing parameters3.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2022