Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29595 |
| Brand Name: | CEREBRO Specimen Tracking System |
| Model: | 21.2793, 21.2201, 49.0051, S21.4621, 26.0001, 26.0005, 26.0003, 26.0008, 45.0001, 45.0005, 65.6400 |
| Affected: | on the customer list |
| Software version: | |
| Recalling Organisation: | Leica BioSystems Melbourne Pty Ltd 495 Blackburn Road Mt Waverley VIC 3149 Australia |
| Contact Information: | 021 0841 7323 |
| Manufacturer: | Leica BioSystems Melbourne Pty Ltd |
| Issue: | The manufacturer is aware of seven vulnerabilities (collectively known as Access:7) associated with the remote access solution (Axeda Agent and Axeda Desktop Server for Windows) produced by Axeda. It has identified that the Axeda vulnerabilities impact Leica Biosystems CEREBRO System with the RemoteCare solution installed. The Axeda software and Leica RemoteCare for CEREBRO was phased out at the end of 2020. RemoteCare and is no longer distributed or supported. However, the CEREBRO System could be vulnerable if the RemoteCare software remains installed. Customers with RemoteCare software still installed on the CEREBRO System should contact a local Sales and Support Representative for assistance with removal of the software. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/6/2022 |