Detail of Recall

Type of Product:Device
Medsafe Reference:29595
Brand Name:CEREBRO Specimen Tracking System
Model:21.2793, 21.2201, 49.0051, S21.4621, 26.0001, 26.0005, 26.0003, 26.0008, 45.0001, 45.0005, 65.6400
Affected:on the customer list
Software version:
Recalling Organisation:Leica BioSystems Melbourne Pty Ltd
495 Blackburn Road
Mt Waverley
VIC 3149
Australia
Contact Information:021 0841 7323
Manufacturer:Leica BioSystems Melbourne Pty Ltd
Issue:The manufacturer is aware of seven vulnerabilities (collectively known as Access:7) associated with the remote access solution (Axeda Agent and Axeda Desktop Server for Windows) produced by Axeda.
It has identified that the Axeda vulnerabilities impact Leica Biosystems CEREBRO System with the RemoteCare solution installed. The Axeda software and Leica RemoteCare for CEREBRO was phased out at the end of 2020. RemoteCare and is no longer distributed or supported. However, the CEREBRO System could be vulnerable if the RemoteCare software remains installed.
Customers with RemoteCare software still installed on the CEREBRO System should contact a local Sales and Support Representative for assistance with removal of the software.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/6/2022