Detail of Recall

Type of Product:Device
Medsafe Reference:29594
Brand Name:V680 Ventilator
Model:850011: Philips V680 Ventilator
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics California LLC
Issue:The manufacturer has identified that during dual-limb invasive ventilation, a patient cough of sufficient magnitude to drive circuit pressures above 95 cmH2O for longer than 150 milliseconds may cause a "Vent Inoperative 1008: Machine and Proximal Pressure Sensors Failed" alarm, which will cause the V680 ventilator to cease therapy and the ventilator will not function, however, it will remain powered on. 22
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory