Detail of Recall

Type of Product:Device
Medsafe Reference:29588
Brand Name:Cholestech LDX® Lipid Profile GLU Cassette and Cholestech LDX® TC HDL GLU Cassette
Model:
Affected:Catalogue Numbers: 10-991 and 10-990 - All lots
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd - Rapid Diagnostics Divison
Ground Floor, Bldg. D
4 Pacific Rise
Mount Wellington
AUCKLAND 1060
Contact Information:0800 466 951
Manufacturer:Alere San Diego Inc
Issue:The manufacturer has identified that bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can impact the accuracy of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/6/2022