Detail of Recall

Type of Product:Device
Medsafe Reference:29561
Brand Name:Merlin Patient Care System Model 3650
Model:Software Model 3330 25.3.X or earlier when used with Accent™/Anthem™ and Endurity™/Assurity™/Allure™
Affected:
Software version:
Recalling Organisation:Abbott Medical New Zealand Limited
4 Pacific Rise
Mt Wellingtion
Auckland 1640
Contact Information:0800 756 269
Manufacturer:St Jude Medical (USA)
Issue:The manufacturer is notifying customers of the potential for programmer software (MerlinTM PCS) and remote monitoring software applications (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers1. Pacemaker/battery functionality, therapy delivery, and longevity remain normal and within specifications. Voltage measurements and ERI (elective replacement indicator), which is based on direct voltage measurement, remain accurate.
The manufacturer has developed a software update to improve accuracy of predicted battery longevity. Remote monitoring (Merlin.net) and programmer software updates are targeted to begin June 2022 based on geography.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/6/2022