Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29558 |
| Brand Name: | Anaesthesia Work Stations |
| Model: | Avance CS2 and Avance CS2 Pro Anesthesia Systems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | The manufacturer has identified that the Avance CS2, Avance CS2 Pro, and Aisys CS2 Anesthesia Systems base can have a crack in a specific location in the rear of the anesthesia system. If excessive load is applied to the anesthesia device (for example when the device is being moved over a threshold), a cracked base could potentially fracture, resulting in a tip or overbalance of the anesthesia system, which can result in potential injury if it falls on a person. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/6/2022 |