Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29532 |
| Brand Name: | Anti-IA2 ELISA (IgG) |
| Model: | Product no. EA 1023-9601 G |
| Affected: | E220215BA |
| Software version: | |
| Recalling Organisation: | ESL Biosciences (NZ) Ltd 188 St Heliers Bay Road St Heliers AUCKLAND 1071 |
| Contact Information: | |
| Manufacturer: | Euroimmun Medizinische Labordiagnostika AG |
| Issue: | Due to customer complaints in relation to invalid positive controls due to low optical density (OD) values for calibrator 3 during the processing of the Anti-IA2 ELISA (IgG) (order no. EA 1023-9601 G) of lot E220215BA, we have initiated an investigation to confirm the existence of the problem, identify its cause and determine an appropriate corrective measure. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/6/2022 |