Detail of Recall

Type of Product:Device
Medsafe Reference:29526
Brand Name:MITG - Palindrome Chronic Catheter Kits
Model:
Affected:Affected lots will start with 17144xxxxx through 22038xxxxx. Example: 1714400164. If your Lot number is outside this range, your product is not impacted.
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Brand Name:MITG - RCS - Palindrome Precision Chronic Catheter Kit with VenaTrac - Catheter, haemodialysis, implantable
Model:
Affected:Affected lots will start with 17144xxxxx through 22038xxxxx. Example: 1714400164. If your Lot number is outside this range, your product is not impacted.
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Issue:The manufacturer has identified a potential leaking condition within the hub of specific chronic hemodialysis catheters. Flushing one extension tube may result in unanticipated fluid return through the adjacent extension tube (in addition to the anticipated flow of fluid through the distal tip of the catheter). The condition is the result of an inter-lumen void formed in the hub component of the catheter during the manufacturing process. During use, this observed condition could translate to cross-communication of the blood circuit.
To date, Medtronic has received one reported complaint and has not received any reports of patient injury or death related to this issue.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2022