Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29501 |
| Brand Name: | Leica CM1950 and Leica CM3050 S |
| Model: | Leica CM1950: All devices in serial number range: 9492 - 10709. Leica CM3050 S: All devices in serial number range: 7825 - 8062 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Leica Microsystems Pty Ltd, New Zealand Branch 3/33 Spartan Road Takanini Auckland 2105 |
| Contact Information: | 021 0841 7323 |
| Manufacturer: | Leica Biosystems Nussloch GmbH |
| Issue: | The manufacturer has identified that the cooling valve design may not be sufficient for all uses, as it is acknowledged, that use cases can vary. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/6/2022 |