Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29498 |
| Brand Name: | Citadel Floatation Therapy System |
| Model: | Arjo Citadel Patient Care Systems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Arjo New Zealand Limited 34 Vestey Drive Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 000 151 |
| Manufacturer: | ArjoHuntleigh AB |
| Issue: | The manfacturer has observed that the e-bay box might overheat while the device is plugged into mains. This overheating can happen when the grounding screw inside the e-bay boxis touching the inner side ofthe e-bay box cover.This can lead then to electric circuit closure ande-bay box overheating |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/6/2022 |