Type of Product:Device
Medsafe Reference:29470
Brand Name:Alinity m System
Model:08N53-002
Affected:
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:
Issue:The manufacturer has received complaints associated with the Alinity m System where customers reported experiencing an issue referred to as "infinite solution". It was observed that system solution (usually Alinity m Lysis Solution) was being transferred from the reagent cradle bottle into the system solution's reservoir through the lines without a transfer command initiated, resulting in reagent transfer errors and instrument stoppage.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/5/2022