Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29470 |
| Brand Name: | Alinity m System |
| Model: | 08N53-002 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Abbott Australasia Pty Ltd (NZ) Abbott Laboratories NZ 4 Pacific Rise, Mt Wellington Auckland 1060 |
| Contact Information: | 61 2 93849717 |
| Manufacturer: | |
| Issue: | The manufacturer has received complaints associated with the Alinity m System where customers reported experiencing an issue referred to as "infinite solution". It was observed that system solution (usually Alinity m Lysis Solution) was being transferred from the reagent cradle bottle into the system solution's reservoir through the lines without a transfer command initiated, resulting in reagent transfer errors and instrument stoppage. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/5/2022 |