Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29457 |
| Brand Name: | INFUSION PUMP |
| Model: | All MiniMed™ 600 series and MiniMed™ 700 series insulin pump models, including MiniMed™ 640G |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic MiniMed |
| Brand Name: | DIB - 670G |
| Model: | All MiniMed™ 600 series and MiniMed™ 700 series insulin pump models, including MiniMed™ 640G, MiniMed™ 670G, MiniMed™ 700, MiniMed™ , MiniMed™ 770G, and MiniMed™ 780G |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic MiniMed |
| Brand Name: | 700G series Pumps |
| Model: | Model: All MiniMed™ 600 series and MiniMed™ 700 series insulin pump models, including MiniMed™ 640G, MiniMed™ 670G, MiniMed™ 700, MiniMed™ , MiniMed™ 770G, and MiniMed™ 780G |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic MiniMed |
| Brand Name: | DIA - 780G pump |
| Model: | Model: All MiniMed™ 600 series and MiniMed™ 700 series insulin pump models, including MiniMed™ 640G, MiniMed™ 670G, MiniMed™ 700, MiniMed™ , MiniMed™ 770G, and MiniMed™ 780G |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic MiniMed |
| Issue: | Update 20/05/24:The manufacturer has previously reached out to patients about a potential issue with the battery cap on the MiniMed™ 600 series and MiniMed™ 700 series insulin pumps and provided actions they should take. The manufacturer has now have developed a new battery cap for these pumps, which addresses the potential issues with the previous battery cap (model ACC-1527). The battery cap on the pump consists of a plastic cap and a metal contact that work together with the AA battery to power the pump. The metal contact should be held in place by three raised, round, black, plastic dots. If the metal contact becomes loose or falls off from the battery cap, it can result in an incomplete battery connection, leading to no power source to the pump. When the pump detects no power source, an "Insert battery" alarm will occur, and insulin delivery will immediately stop. After 10 minutes, the alarm sound may increase to a siren, and the pump will turn off. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/5/2024 |