Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29425
Brand Name:Liko Overhead Lifting hoists & components
Model:Any UNIVERSAL TWINBAR 670 QRH™ manufactured before 28OCT2021. 3156087 & P3156087
Affected:
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Liko AB
Issue:The manufacturer has determined that there is a potential for patient fall with the Universal Twinbar 670 QRH™, which may have been manufactured with an incorrect, shorter center bolt. The potentially hazardous situation can occur when the Twinbar with the shorter center bolt, is used in combination with a Liko® Scale and a rigid (non-rotating) interface.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/9/2022