| Type of Product: | Device |
| Medsafe Reference: | 29425 |
| Brand Name: | Liko Overhead Lifting hoists & components |
| Model: | Any UNIVERSAL TWINBAR 670 QRH™ manufactured before 28OCT2021. 3156087 & P3156087 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Liko AB |
| Issue: | The manufacturer has determined that there is a potential for patient fall with the Universal Twinbar 670 QRH™, which may have been manufactured with an incorrect, shorter center bolt. The potentially hazardous situation can occur when the Twinbar with the shorter center bolt, is used in combination with a Liko® Scale and a rigid (non-rotating) interface. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/9/2022 |