Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29409 |
| Brand Name: | Patient monitor, central unit |
| Model: | CARESCAPE Central Station (CSCS) V2 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | The manufacturer has become aware of an issue with the CARESCAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station. Patient monitoring at the bedside is not affected. Loss of monitoring can result in delayed response to a change in patient clinical condition. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/5/2022 |