Detail of Recall

Type of Product:Device
Medsafe Reference:29408
Brand Name:Avanos Australia - Cortrack 2 Enteral Access System - Enteral tube tracking system
Model:Product Code 20-0950
Affected:
Software version:
Recalling Organisation:Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics
54 Carbine Road
Mt Wellington
AUCKLAND
Contact Information:09 969 0705
Manufacturer:Avanos Medical Inc
Brand Name:Halyard Australia - Cortrack 2 Enteral Access System - Enteral tube tracking system
Model:Product Code 20-0950
Affected:
Software version:
Recalling Organisation:Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics
54 Carbine Road
Mt Wellington
AUCKLAND
Contact Information:09 969 0705
Manufacturer:Halyard Health Inc
Issue:The manufacturer is making updates to the labeling in the Operator's Manual and training documents stating that the confirmation of NG/NI tubes placed using CORTRAK 2 should be confirmed per institutional protocol.
This incorrect positioning has resulted in reports of 60 injuries and 23 patient deaths in the United States and United Kingdom, while using the CORTRAK* 2 Enteral Access System, since 2015.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/5/2022