Detail of Recall

Type of Product:Device
Medsafe Reference:29384
Brand Name:MITG - Trocar sleeve (Class IIA)
Model:EEA25, EEAXL25, EEA2535, EEAXL2535
Affected:All Lots beginning with "P0", "P1", "P7", "P8", "P9"
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Issue:Distributed 25mm EEA Autosuture Circular Staplers with DST Series Technology, with model numbers EEA25, EEAXL25, EEA2535, and EEAXL2535, have the potential for the staple guide to not be securely attached to the instrument. This issue is related only to the 25mm EEA Autosuture Circular Staplers with DST Series Technology. No other Medtronic products or other sizes of EEA Autosuture Circular Staplers with DST Series Technology are affected by this issue.
Through 04 April 2022, Medtronic has received 23 complaints potentially related to disengaged staple guides, of which two (2) complaints were directly confirmed through investigation of returned product. A staple guide not attached to the instrument could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples. This could result in delay of treatment, extended hospital stay, unspecified tissue injury, unintended radiation exposure, unexpected medical intervention and others.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/5/2022