Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29354 |
| Brand Name: | Steris Washers/Disinfectors |
| Model: | ST-FH03-062 SYNERGY WASHER 415V PASS-THROUGH - STEAM, ST-FH03-063 SYNERGY WASHER 415V PASS-THROUGH - ELECTRIC |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Device Technologies New Zealand 47 Arrenway Drive Albany AUCKLAND 0632 |
| Contact Information: | 09 913 2000 |
| Manufacturer: | Steris Canada Corporation |
| Issue: | The manufacturer has identified that the electrical contactor component present in the drying chamber of the Reliance Synergy Washer/Disinfector could malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/5/2022 |