Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29349 |
| Brand Name: | 9100E-Series Washer Disinfector |
| Model: | 9125E-001, 9128E-001 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Getinge Disinfection AB |
| Issue: | The manufacturer has discovered that verification of installation is not always documented or the documentation is incomplete and does not confirm that all installation activities have been performed as required by the installation manual. 1. In the final testing, the device is not tested in the final assembly, as for example the doors are to heavy and will be assembled only onsite at the customer site. 2. The installation manual is however missing a protocol to be used by the technician performing the installation to document the performed installation points. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/5/2022 |