Detail of Recall

Type of Product:Device
Medsafe Reference:29349
Brand Name:9100E-Series Washer Disinfector
Model:9125E-001, 9128E-001
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Getinge Disinfection AB
Issue:The manufacturer has discovered that verification of installation is not always documented or the documentation is incomplete and does not confirm that all installation activities have been performed as required by the installation manual.
1. In the final testing, the device is not tested in the final assembly, as for example the doors are to heavy and will be assembled only onsite at the customer site.
2. The installation manual is however missing a protocol to be used by the technician performing the installation to document the performed installation points.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/5/2022