Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29335
Brand Name:Illumina_IVD_NextSeq 550Dx Instrument (20005715)
Model:MiSeqDx Instrument
Affected:Catalog Number DX-410-1001
Software version:
Recalling Organisation:Clinical & Regulatory Services Limited
30 Fulford Road
Havelock North
Hastings 4130
Contact Information:(06) 844 4490
Manufacturer:Illumina Inc
Issue:The manufacturer has released a new software version that remediates the cybersecurity vulnerability and restores remote access to the LRM Web User Interface. New versions of the NextSeq Operating Software (NOS) v1.7.0 and NextSeq Control Software (NCS) v4.2.0, which both include Local Run Manager (v4.1.0 and v4.0.0, respectively), are available for installation on the NextSeq 550Dx instruments.
Update 5/10 2023: In May 2022, Illumina contacted customers regarding a Local Run Manager (LRM) cybersecurity vulnerability which included the NextSeq™ 550Dx instruments. The manufacturer is following up regarding the Local Run Manager (LRM) cybersecurity vulnerability. This information outlines the issue summary, Illumina's actions, and required customer actions.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/10/2023