Detail of Recall

Type of Product:Device
Medsafe Reference:29307
Brand Name:Ventilators, continuous, life supporting
Model:CARESCAPE R860
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufactuer has recommended replacement backup batteries manufactured on or after April 1, 2019, for CARESCAPE R860, Engström Carestation, and Engström PRO can fail earlier than their estimated life. For these batteries, the alarm that alerts the user on battery run time remaining could potentially be inaccurate. This issue could result in the ventilator shutting down sooner than indicated by the alarm when running on the backup battery, which could potentially lead to loss of ventilation.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/4/2022