Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29299 |
| Brand Name: | Sensis / Sensis Lite |
| Model: | |
| Affected: | |
| Software version: | AXIOM Sensis XP or Sensis / Sensis Lite with VC12M software |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | Multiple software issues on AXIOM Sensis XP or Sensis / Sensis Lite with VC12M. Issue 1 - Software crash during split screen Issue 2 - Software crash due to system internal timeout Issue 3 - Missing regular reboot of the system may result in frozen vital signs Potentially high risk as this can potentially result in an incorrect treatment or diagnosis. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/4/2022 |