Detail of Recall

Type of Product:Device
Medsafe Reference:29293
Brand Name:Anaesthesia System - Getinge Flow - C
Model:PART NUMBER 6887700
Affected:4742, 4132, 4930, 4931, 4932, 4933, 4934 and 4935
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Critical Care AB
Issue:The suction unit on/off knob material on the Getinge Flow-c Getinge anesthesia systems have a tendency to crack as shown in the pictures below. The issue has not been identified to begin at any specific time, or with any particular batch, therefore all units that are installed may be affected.
In worst case the knob can break and then the suction unit cannot be activated. If it breaks in active mode the vacuum regulator knob can still be used to regulate the pressure of the suction.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/4/2022