Detail of Recall

Type of Product:Device
Medsafe Reference:29275
Brand Name:ConMed Endoscopic Diathermy Electrodes
Model:item code: 60-5274-132
Affected:From 20190801X to 20211031X
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:ConMed Corporation
Issue:The manufacturer received reports that the tip of the electrode could detach during use. In most cases, the tip was easily retrieved by the surgical team during the surgery. If the tip detaches into the patient's cavity and cannot be retrieved, there should be little to no potential consequences in leaving the electrode tip in the patient as the electrode tip is made of stainless steel. The surgeon should make a decision, based on the surgeon's own clinical experience and the patient's individual needs, regarding the manner and time spent in attempting to retrieve the electrode tip.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/4/2022