Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29255 |
| Brand Name: | Siemens Artis Q, Q.zen, zeego, zee systems |
| Model: | Catalogue number - 11327600, 11327700 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has identified that in some rare cases an interruption of the protective earth and loss of detection of fault current might occur. ARTIS icono biplane, ARTIS icono floor - this results in Failure in the routing of the protective earth |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/4/2022 |