Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29255
Brand Name:Siemens Artis Q, Q.zen, zeego, zee systems
Model:Catalogue number - 11327600, 11327700
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has identified that in some rare cases an interruption of the protective earth and loss of detection of fault current might occur.
ARTIS icono biplane, ARTIS icono floor - this results in Failure in the routing of the protective earth
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/4/2022