Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29253 |
| Brand Name: | VITEK® 2 GP ID Test Kit |
| Model: | |
| Affected: | 21342 |
| Software version: | |
| Recalling Organisation: | Medi'Ray New Zealand Ltd 53-55 Paul Matthews Road Albany Auckland 0632 |
| Contact Information: | 0800 633 472 |
| Manufacturer: | BioMerieux New Zealand Ltd |
| Issue: | The manufacturer has identified a performance issue associated with the SAC (SACCHAROSE/SUCROSE) reaction during testing of Listeria monocytogenes BAA-751 and the VITEK 2 GP ID test kit. The BAA-751 strain is referenced in the GP ID Package Insert for QC testing. Investigation activities concluded that Listeria monocytogenes BAA-751 may not provide a reliable negative reaction for the SAC well. Incorrect QC results can potentially lead to a delay in the release of patient results. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/4/2022 |