Detail of Recall

Type of Product:Device
Medsafe Reference:29250
Brand Name:Floseal Applicators_47187_BS Zurich
Model:Product Code: ADS201865
Affected:EASYGRIP FLO-41, lot 0000357425 and 0000366117
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare Corporation SA
Issue:The maufacturer is issuing an Urgent Medical Device Recall for one lot of EASYGRIP FLO-41 due to the potential use of product after their expiry date. The human-readable expiry dates printed on the primary and secondary package labeling are in the correct YYYY-MM-DD format, how the manufacturer identified that the expiry date encoded in the 2D barcode on this labeling is in YY-DD-MM format. If only the 2D barcode is used to identify the product expiry date, and the user is under the impression the date is in YY-MM-DD format, there is potential for these lots to be used past the actual expiry date.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/4/2022