Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29191 |
| Brand Name: | Bothell ECR External/Adult Pads |
| Model: | Philips M5071A (adult) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Brand Name: | HeartStart Infant/Child-Smart Pads Cartridge |
| Model: | M5072A (infant/child) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | The manufacturer has become aware that the HS1/OnSite/Home AED pads (PN: M5071A, M5072A) have experience gel separation from the foam/tin backing when peeled from the yellow plastic liner. The gel may fold onto itself resulting in reduced surface area of gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad. Any pad currently installed in or stored with an HS1/OnSite/Home AED could experience this problem, and it is not possible to know prior to patient use if your pad is affected because the pads are protected by a foil seal. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Public Recall |
| Date Commenced: | 10/1/2023 |