Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29191
Brand Name:Bothell ECR External/Adult Pads
Model:Philips M5071A (adult)
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Brand Name:HeartStart Infant/Child-Smart Pads Cartridge
Model:M5072A (infant/child)
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has become aware that the HS1/OnSite/Home AED pads (PN: M5071A, M5072A) have experience gel separation from the foam/tin backing when peeled from the yellow plastic liner. The gel may fold onto itself resulting in reduced surface area of gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad. Any pad currently installed in or stored with an HS1/OnSite/Home AED could experience this problem, and it is not possible to know prior to patient use if your pad is affected because the pads are protected by a foil seal.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Public Recall
Date Commenced:10/1/2023
Hide menus