| Type of Product: | Device |
| Medsafe Reference: | 29178 |
| Brand Name: | PrisMax |
| Model: | |
| Affected: | PrisMax System - product code 955558 - All serial numbers, PrisMax System - product code 955725 - All serial numbers |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare SA (Switzerland) |
| Issue: | The manufacturer has identified a software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/3/2022 |