Detail of Recall

Type of Product:Device
Medsafe Reference:29178
Brand Name:PrisMax
Model:
Affected:PrisMax System - product code 955558 - All serial numbers, PrisMax System - product code 955725 - All serial numbers
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare SA (Switzerland)
Issue:The manufacturer has identified a software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/3/2022