Type of Product:Device
Medsafe Reference:29172
Brand Name:Angiographic X-Ray
Model:Azurion 3 M12 722063, Azurion 7 B12 722067, Azurion 3 M15 722064 Azurion 7 B20 722068, Allura Xper R9 7 M12 722065 Azurion 7 M12 722078, Allura Xper R9 7 M20 722066 Azurion 7 M20 722079
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has become aware that the Azurion system, the user can add a new study to a patient by selecting the option Add Study. The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure, the Patient Type changes inadvertently to a Patient Type different than the one selected
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/3/2022