Type of Product:Device
Medsafe Reference:29155
Brand Name:Olympus - Lumen extraction balloon, single-use (Japan)
Model:FG-51D, FG-55D
Affected:71K, 81K, 91K, 01K, 11K, 72K,82K,92K, 02K,12K,73K,83K,93K,03K,13K,74K,84K,94K,04K,14K,75K, 85K,95K,05K,15K,76K,86K,96K,06K,16K,77K,87K,97K,07K,17K,78K,88K,98K,08K,18K,79K,89K,99K,09K,19K,7XK,8XK,9XK,0XK,1XK,7YK,8YK,9YK,0YK,11H,7ZK,8ZK,9ZK, 0ZK,77H
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer has received complaints that the grasping portion of FG-51D forceps can be difficult to open and close. This may result in the following hazards:
· The user notices unusual stiffness during inspection before use. It may cause the prolongation of the procedure.
· If the user pushed with extra force, the grasping portion (basket) suddenly extruding beyond the sheath and it pushes the mucosa, which may cause mucosal injury.
· Difficult to perform procedure. It may cause the prolongation of the procedure.
· Device replacing. It may cause the prolongation of the procedure.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/3/2022