Detail of Recall

Type of Product:Device
Medsafe Reference:29154
Brand Name:VAS - InPact Admiral Paclitaxel-eluting PTA Balloon Catheter
Model:Model: SBI06012013S - product code - SBI06012013S
Affected:0010875709
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:During a routine inspection, the manufacturer noted that there was damage to the sterile pouch that contains the IN.PACT Admiral catheter. An investigation determined that there had been a change implemented to one manufacturing line which may cause pouch damage. All batches manufactured on this line after that change are being retrieved.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/3/2022