Type of Product:Device
Medsafe Reference:29075
Brand Name:Bellavista Ventilators
Model:generation 6 (G6) hardware
Affected:
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Imtmedical AG
Issue:The manufacturer has received reports that on bellavista ventilators with generation 6 (G6) hardware, ventilation may cease during clinical use, and ventilation monitoring waveforms and parameters may become frozen (not updated). The ventilator generates a continuous audible and visual high priority alarm for technical failure 305. Ventilation is suspended until the unit is rebooted or replaced.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/3/2022