Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 29075 |
| Brand Name: | Bellavista Ventilators |
| Model: | generation 6 (G6) hardware |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | Imtmedical AG |
| Issue: | The manufacturer has received reports that on bellavista ventilators with generation 6 (G6) hardware, ventilation may cease during clinical use, and ventilation monitoring waveforms and parameters may become frozen (not updated). The ventilator generates a continuous audible and visual high priority alarm for technical failure 305. Ventilation is suspended until the unit is rebooted or replaced. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/3/2022 |