Type of Product:Device
Medsafe Reference:29060
Brand Name:Hemodynamic Application
Model:Philips Hemodynamic Application R1.2.2 and R1.2.3
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified that when a Neonatal patient is admitted in the Philips Hemodynamic Application, the Non-Invasive Blood Pressure is not set to Neonatal on the Philips Xperflex Cardio FC2010 patient monitor. The Philips Hemodynamic Application will indicate 'Neonatal' on the Non-Invasive Blood Pressure (NIBP) field displayed on the User Interphase (UI) in the control and exam room, but the FC2010 NIPB will be set in Adult mode.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/3/2022