| Type of Product: | Device |
| Medsafe Reference: | 29039 |
| Brand Name: | Stress Exercise monitoring system - Cardiac |
| Model: | |
| Affected: | XScribe – 5.xx to 6.xx (all versions from 5.01 to 6.3.1) |
| Software version: | |
| Recalling Organisation: | Ebos Group Ltd 108 Wrights Road Addington Christchurch |
| Contact Information: | 0800 18 17 16 |
| Manufacturer: | Welch Allyn Inc |
| Brand Name: | Electrocardiograph, Holter analyser, software |
| Model: | |
| Affected: | HScribe – 5.xx and 6.x.x (all versions from 5.01 to 6.4.0) |
| Software version: | |
| Recalling Organisation: | Ebos Group Ltd 108 Wrights Road Addington Christchurch |
| Contact Information: | 0800 18 17 16 |
| Manufacturer: | Welch Allyn Inc |
| Brand Name: | Interpretive multichannel electrocardiograph |
| Model: | |
| Affected: | RScribe 5.xx, 6.x.x, and 7.x.x (all versions from 5.01 to 7.0.0) |
| Software version: | |
| Recalling Organisation: | Ebos Group Ltd 108 Wrights Road Addington Christchurch |
| Contact Information: | 0800 18 17 16 |
| Manufacturer: | Welch Allyn Inc |
| Issue: | The manufacturer has become aware of a software vulnerability, which allows the entry of any username, that is provisioned in the application, without supplying a password. Entry of such username will grant access, with the same privileges as the username, to the software application within the medical device. This vulnerability occurs only when the device is activated with the Single Sign-On (SSO) feature in either stand-alone or client/server configurations. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/8/2022 |