Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:29019
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:
Affected:Catalogue number – 10849000, 11327600, 11327700 and 11327667
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has indentified a potential issue with Artis system in syngo Application Software version VE20. After CT image data from Toshiba is loaded, unintended image mirroring can occur along the horizontal and vertical image axes. The patient orientation/position may be misinterpreted.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/2/2022