Detail of Recall

Type of Product:Device
Medsafe Reference:29015
Brand Name:Alinity m STI Amp Kit
Model:
Affected:517789, 518275, 519550, 519599, 520273, 520401
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Bioresearch Centre
Issue: There is potential for incorrect results (false negative Neisseria gonorrhoeae) results in the following scenario:
For customers that run the 09N17-091 kit with the unintended Alinity m STI Application Specification File (09N17-01B), any Neisseria gonorrhoeae (NG) results with a Cycle Number (CN) between 40.00 and 41.00 would be called Negative. Those results would be called Positive with the Alinity m STI Application Specification File (09N17-05A).
Therefore, there is a potential risk for False Negative results for NG samples with a CN between 40.00 and 41.00.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/2/2022