Detail of Recall

Type of Product:Device
Medsafe Reference:28975
Brand Name:Atellica CH 930 Analyzer
Model:
Affected:Affected product - Atellica CH Calcium_2 All lots, Atellica CH Enzymatic All lots, Atellica CH Fructosamine all lots, Atellica CH Lithium_2 All lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has received multiple complaints reporting issues with carryover from several assays. Per Design Input Requirements (DIR), carryover must not be = 5% interference or ± one significant figure unit increment (example: 0.1 mg/dL compared to 0.2 mg/dL).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/2/2022