Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28951 |
| Brand Name: | Pulsion PCCO Monitoring Kit |
| Model: | |
| Affected: | Lots - 21BB01 |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Pulsion Medical Systems SE |
| Issue: | The manufacturer has identified that the listed products have been produced with packaging material that is suspected to have an uncoated area. The absence of coating on the inner material surface leads to a weak seal between blister and lid of the sterile packaging. Due to the vacuum process within ETO-sterilization or transportation, the lid will be detached from the blister. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/2/2022 |