Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28951
Brand Name:Pulsion PCCO Monitoring Kit
Model:
Affected:Lots - 21BB01
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Pulsion Medical Systems SE
Issue:The manufacturer has identified that the listed products have been produced with packaging material that is suspected to have an uncoated area. The absence of coating on the inner material surface leads to a weak seal between blister and lid of the sterile packaging. Due to the vacuum process within ETO-sterilization or transportation, the lid will be detached from the blister.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/2/2022