Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28933 |
| Brand Name: | Ysio Max |
| Model: | Ysio fully automated systems with VC10 - |
| Affected: | Catalogue number – 10281013 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has become aware that in another software version of the Ysio fully automated system (VB10) it was identified that the system may incorrectly process the position values of the ceiling tube stand. If the user activates a longitudinal or transversal stand movement at the time of this error, the system controls may initiate a correction of the tube stand position, resulting in an unexpected movement. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/2/2022 |