Detail of Recall

Type of Product:Device
Medsafe Reference:28933
Brand Name:Ysio Max
Model:Ysio fully automated systems with VC10 -
Affected:Catalogue number – 10281013
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has become aware that in another software version of the Ysio fully automated system (VB10) it was identified that the system may incorrectly process the position values of the ceiling tube stand. If the user activates a longitudinal or transversal stand movement at the time of this error, the system controls may initiate a correction of the tube stand position, resulting in an unexpected movement.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/2/2022